Display for reference force transducers model INDI 00

Display for reference force transducers model INDI 00
The INDI-00 is a high accuracy portable indicator designed to be used in combination with standard reference force transducers with accuracy class 0.5 and 00 following the norm ISO376.
  • Display 6 digits (red leds of 15 mm)
  • Units Newton, kilo Newton or mV/V
  • Input Wheatstone bridge (6 wires - 50ohms min.)
  • Maximum 4 mV/V
  • Transducer supply 5 Volt (inverted polarity - 29 Hz)
  • Accuracy class 0.01 %
  • Non-linearity <= 0.002 % of full scale
  • Input resolution 0.1 V/V per digit
  • Internal resolution 550.000 divisions
  • Conversion speed 17 / s
  • Thermal stability of zero <= 2 ppm/°C
  • Thermal stability of span <= 2 ppm/°C
  • Long term stability <= 0.005 % of full scale per year
  • Digital outputs RS-232C or (RS-485 on option)
  • Analog output (option) 0..10 Volt or 4..20 mA (16 bits - drain)
  • 2 independant set points (option) 24 VDC - 100 mA (transistor output)
  • Power supply 230 VAC (option:internal rechargeable battery)
  • Operating temperature from -10 to +40°C
  • Storage temperature from -10 to +70°C
  • Humidity 90 % RH
  • Protection IP 54
  • Internal calibration resistor activated by push button on the front plate
  • Weight 6.7 kg
  • Acquisition and management software (see specific documentation)